• UA
  • EN
  • About Us
    • History
    • Company Profile
    • Certificates&Patents
  • Products
    • Soneus P7
    • Soneus P5
    • Soneus P3
    • OEM Products
    • BFS blood flow simulator
    • Services
  • Events
  • Support
    • Recommendations
    • Articles
    • Feedback
    • Warranty
  • Contacts
Joomla

BFS blood flow simulator

BFS Doppler electronic blood flow simulator.

 

Metrologist and Manufacturer Assistant


Effective Tool of testing of ultrasound Doppler diagnostic Simulators and ultrasound systems in Doppler mode.
To purchase the product, please call ULTRASIGN TECH Company's specialists  +38 067 359 2662.
 
  • Purpose
  • Specification
  • Delivery set

Purpose

BFS Doppler electronic blood flow simulator is designed for periodic verification, verification of parameters and operation of ultrasound Doppler diagnostic Simulators and ultrasound systems in Doppler mode, as well as spectral and color modes of blood flow mapping.

The simulator is used in regulation, acceptance and other tests of ultrasound Doppler diagnostic devices and systems, as well as for their initial and periodic verification in operation. The simulator can be used for testing ultrasound Doppler diagnostic devices and systems according to IEC TS 61895 and CEI/IEC 61206 in the part that provides or allows the use of electronic blood flow simulator.

Suitable for use by metrological services and manufacturers of ultrasound Doppler diagnostic simulators and ultrasound systems with Doppler modes.

Specification

The simulator is made in a portable version powered by ~ 220 V.

Technical data and characteristics:

Name of characteristic

Value

nominal value range
1 2 3
AC supply voltage, V 220 100 - 250
Frequency, Hz 50 47 - 65
Operating acoustic frequency range, MHz - from 2 to 7,5
Repeat frequency range (RPF) of the verified simulators probing pulse, kHz - from 1 to 25
Range of duration of the verified simulators probing pulses, μs - from 0,12 to 4
Simulation signal readiness time after receiving the request signal, s no more 5  
Relative instability of the master quartz oscillator - ± 50×10-6

Forming simulation signal modes:

- by the form of radiation

- by the spectrum of the Doppler signal

 

pulsed and continuous;

harmonic, biharmonic and noise-like with stable frequencies or pulse simulation

Continuous operation time

not limited
BFS block dimensions, mm 200x135x75

Metrological characteristics:

Name of parameter Value
1 2
Doppler frequency shift control range, kHz from minus 45 to 45
Non-uniformity of amplitude-frequency characteristic in the Doppler range, dB no more 1
Mirror component level, dB no more minus 60
Adjustment range of the output level (at harmonic Doppler signal), dB no less 60
Limiting error of the output level range from 0 dB to -60 dB, dB no more 1
Pulsation frequency adjustment range, Hz from 0,2 to 50
Limiting relative error of pulsation frequency setting no more ± 1×10-4
Limiting relative error δf of forming of the maximum frequency of the broadband Doppler signals spectrum no more ± 0,5×10-2
Limiting relative error of forming of the average frequency of the broadband Doppler signals spectrum no more ± 0,3×10-2

Delivery set

  • BFS block
  • Laptop with software and power supply
  • Power supply
  • User manual
  • Passport
  • Bag


 
ultrasign TECH lLC
Address: 46, Chernyishevskaya Str., Kharkov, 61002, Ukraine
Office: 120 E10, Academica Pavlova Str., office 411, Kharkov, 61054, Ukraine
Tel: +38 067 359 2662
Website: ultrasign.com.ua
E-mail: info@ultrasign.com.ua
E-mail: tech@ultrasign.com.ua
ALL RIGHTS RESERVED "ULTRASIGN TECH LLC" 2014-2023
YJSG_LINKS_VALIDCSS | YJSG_LINKS_VALIDHTML | YJSG_LINKS_TOP
YJSG_LINKS_YOUJOOMLA_BRANDING